K-Th10 Side Navigation Ordering & Pricing K-8 Datasheet Flicker Measuring Software Specifications Colorimeter Accessories Ordering & Pricing K-8 Datasheet Flicker Measuring Software Specifications Technical Spec Page Ordering & Pricing K-8 Datasheet Flicker Measuring Ordering & Pricing K-8 Datasheet Flicker Measuring Software Specifications Technical Spec Page Ordering & Pricing K-8 Datasheet Flicker Measuring
Colorimeters: K-10 LF: Technical Specifications

 

Klein K-10 LF Colorimeter Technical Specs (For K-10 SF Technical Specs Click Here)K-8 image

 

 

 

Measuring Function: Colorimetric, xyY, uvY, RGB, Tc Tuv Y,

Measuring Range: 0.00006-2919 fL; 0.0002-10,000 nits (candelas per square meter)

Measurement Area/Distance: 22mm diameter / 110 mm from display

Display Intervals: 0.1 second auto ranging; 6 ranges, each channel ranges independently.

Filtering: 20Hz low pass 3 pole analog filter, 14 bit A/D Converter, digital averaging

System Response: Closely matches CIE 1931 Std Observer curves. Klein vs. CIE.

Accuracy(Illuminant A):
Chroma +- 0.002 x,y (displayed value > 1.00 nits)
Luminance +- 1% Y, +- 1 LSB (displayed value > 0.01 nits)
 
Repeatability:
Chroma +- 0.002 x,y (displayed value > 1.00 nits)
Luminance +- 1% Y, +- 1 LSB (displayed value > 0.01 nits)

Data Interface: 9600 baud RS232 Communication to PC

Output Connectors: Serial com 9-pin (DB9), USB plug for 5v power only

Size: 10.5" Long x 2.375" Diameter (267mm x 60mm)

Power Supply: 5 volts, 75mA nominal (175mA max w/LED's On) from a PCs USB port, or from included power supply

Operating Requirements: Temperature: 0-40 C; Humidity: below 85% RH non-condensation

Weight/Materials: net 19 oz (591g.); Platinum colored high-impact plastic and black anodized aluminum

Shipping Weight: 7 pounds